Hydrogen Ion-Selective Poly(vinyl chloride) Membrane Electrode Based on a Calix[4]arene.

DOI PDF Web Site Web Site Web Site ほか1件をすべて表示 一部だけ表示 被引用文献1件 参考文献72件

この論文をさがす

抄録

A hydrogen ion-selective poly(vinyl chloride) membrane electrode was constructed using 5,11,17,23-tetra-tert-butyl-25,26,27,28-tetracyanomethoxycalix[4]arene as a neutral carrier. The electrode showed an apparent Nernstian response in the 2 - 11.5 pH range with a slope of 54.0±0.2 mV/pH at 20±1°C. This electrode showed a rapid response of the emf to changes in the pH, high ion selectivity with respect to lithium, sodium and potassium, and characteristics similar to those reported for the conventional pH glass membrane electrode. It can be used as a potentiometric indicator electrode in hydrofluoric acid solutions. The effects of iodide, thiocyanate, perchlorate and bromide on the characteristics of the electrode were also considered.

収録刊行物

  • Analytical Sciences

    Analytical Sciences 19 (2), 217-221, 2003

    社団法人 日本分析化学会

被引用文献 (1)*注記

もっと見る

参考文献 (72)*注記

もっと見る

キーワード

詳細情報 詳細情報について

問題の指摘