アルミニウム基金属を用いた中心開放型機械弁の流体動態的特性評価
書誌事項
- タイトル別名
-
- A study on the hydrodynamics of a new bileaflet valve prosthesis using an aluminum alloy.
この論文をさがす
抄録
In this paper we are reporting on a new bileaf let prosthetic heart valve made of an aluminum alloy. Test valves were made to observe characteristics of hydrodynamics around the valves, and to determine the basic valve design. In order to determine the optimal valve design, we had two design parameters which are axis position rate and leaflet curvature. Results of hydrodynamic studies suggested that the optimal combination between axis position rate and leaflet curvature were 70% and 9%, respectively among the test valves. The new bileaflet valve shows almost the same hydrodynamic characteristics as a conventional bileaflet valve.
収録刊行物
-
- 人工臓器
-
人工臓器 21 (2), 620-624, 1992