Characteristics of the Axial-symmetric Transmission Gauge: Deviation and Long Term Stability of Sensitivity Factor
-
- TAKAHASHI Naoki
- ULVAC Inc.
-
- YUYAMA Junpei
- ULVAC Inc.
-
- TUZI Yutaka
- ULVAC Inc.
-
- ARAKAWA Ichiro
- Department of Physics, Gakushuin Univ.
Bibliographic Information
- Other Title
-
- 軸対称透過形電離真空計の感度係数のばらつきと安定性
- ジク タイショウ トウカガタ デンリ シンクウケイ ノ カンド ケイスウ ノ バラツキ ト アンテイセイ
Search this article
Abstract
The deviation of the sensitivity factors for the axial-symmetric transmission gauge (AT gauge) were examined and compared with those for the B-A gauges and the extractor gauges. The calibrations were performed against nitrogen. The deviations for three types of gauges were about ±10%, nearly the same one another. The long-term variation of the sensitivity factor for two AT gauges was also examined. For one sensor, the sensitivity factor decreased 10% during initial two years operation while the factor stabilized within ±3% for the successive three years. The temperature dependence of the sensitivity factor is discussed for the data obtained in the stabilized period.<br>
Journal
-
- Journal of the Vacuum Society of Japan
-
Journal of the Vacuum Society of Japan 51 (6), 377-381, 2008
The Vacuum Society of Japan
- Tweet