462 コリオリ質量流量計における空気混入診断 : 2つのセンサ出力間における相関の揺動による手法(設備の監視,OS-11 システムのモニタリングと診断,総合テーマ「伝統を,未来へ!」)
書誌事項
- タイトル別名
-
- 462 Aerated Flow Diagnosis in Coriolis Mass flowmeter : Method Based on the Fluctuations of Mathematical Function Related to 2 Sensors' Output
抄録
This paper proposes a diagnostic method of an aerated flow that causes significant flow measurement errors in Coriolis mass flowmeter. In order to diagnose the aerated flow at a water line with a Coriolis mass flowmeter, three factors: amplitudes of 2 sensor signals and the phase difference between them-are focused on. Through the experimental investigations, the following facts are derived: 1) Our proposed diagnostic index increases with the increase in the void fraction. 2) The diagnostic index is approximately 0 at the single-phase flow. These results show that this diagnostic index can distinguish the single-phase flow from the aerated flow.
収録刊行物
-
- Dynamics & Design Conference
-
Dynamics & Design Conference 2010 (0), _462-1_-_462-6_, 2010
一般社団法人 日本機械学会