呼気ポート式シングル呼吸回路を用いた従量式換気における機器表示値と測定値の一回換気量比較分析

  • 濱坂 秀一
    国立病院機構医王病院 呼吸器障害・感染症治療研究室 金沢大学大学院 医薬保健学総合研究科保健学専攻
  • 田上 敦朗
    国立病院機構医王病院 呼吸器障害・感染症治療研究室

書誌事項

タイトル別名
  • Comparison of ventilator display values and measured device values for tidal volume in volume-controlled ventilation using an exhalation port type single breathing circuit

説明

<p>In this study, our aim is to quantitatively evaluate and analyze the ventilator display values and the measured device values for tidal volume using three types of respiratory mechanics, for the purpose of conducting a comparative analysis. The artificial ventilators targeted in this study are the Philips Trilogy Evo series and the Breas Medical AB mart Ventilator Vivo 45LS, both of which utilize an exhalation port type single breathing circuit. Measurements were conducted using a measuring instrument and a test lung with adjustable compliance and resistance. As a result, a significant difference was found between the ventilator display values and the measured device values for tidal volume, regardless of the test lung settings. When setting the tidal volume of a ventilator, it is important not to rely exclusively on the displayed values.</p>

収録刊行物

  • 医療機器学

    医療機器学 94 (3), 327-333, 2024

    一般社団法人日本医療機器学会

被引用文献 (1)*注記

もっと見る

参考文献 (3)*注記

もっと見る

詳細情報 詳細情報について

問題の指摘

ページトップへ