水中のジクロロエチレン異性体のガスクロマトグラフィー質量分析法による定量

書誌事項

タイトル別名
  • GC/MS determination of dichloroethylene isomers in water.
  • スイチュウ ノ ジクロロエチレン イセイタイ ノ ガス クロマトグラフィー シ

この論文をさがす

抄録

Capillary GC/MS combined with purge and trap technique has been applied to analysis for traces of dichloroethyleneisomers (DCEs) in water. With the cryo-capillary interface mode, the detection limit was 0.01μg/l with 5ml sample size. Reproducibility ofselected ion monitoring (SIM) chromatographic peak areas were in the range of 4-7%for concentration of 1μg/l, but were 10% for 0.05μg/l concentrations. The SIM peakintensity ratios between monitor ions showed excellent reproducibilities. The resultsobtained with the cryo-column mode were also discussed and compared with those of the cryo-capillary interface mode.

収録刊行物

  • 分析化学

    分析化学 40 (6), T131-T136, 1991

    公益社団法人 日本分析化学会