Comparison of Measured Doses by Film Badge and Other Personnel Dosimeters

  • KAMINAGA Hiroshi
    Division of Health Physics and Safety, Japan Atomic Energy Research Institute
  • FURUTA Toshishiro
    Division of Health Physics and Safety, Japan Atomic Energy Research Institute
  • KUMAZAWA Shigeru
    Division of Health Physics and Safety, Japan Atomic Energy Research Institute
  • TATSUTA Hatsumi
    Division of Health Physics and Safety, Japan Atomic Energy Research Institute Japan Radiation Engineering Co., Ltd

Bibliographic Information

Other Title
  • フィルムバッジと他の個人線量計による被ばく線量測定値の比較

Search this article

Description

In order to study a discrepancy in external dose measured by various types of personnel dosimeters, especially between a film badge and other dosimeters, the energy and directional dependence of 6 kinds of dosimeters was examined both in free air and on the phantom. On the basis of the experimental results, it was found that the backscattered radiation from the body has stronger influences on the dose by a film badge than that by other types of dosimeters and that the discrepancy is mainly caused by the scattered radiation. Therefore, it is suggested that a film badge should be calibrated on the phantom for diminishing the discrepancy.