Detection of Legionella Species in Environmental Water Samples by Loopamp Legionella Detection Reagents Kit

  • SAIKA Takeshi
    Chemotherapy Division, Mitsubishi Kagaku Bio-Clinical Laboratories, Inc.
  • TATEWAKI Mitsuhiro
    Chemotherapy Division, Mitsubishi Kagaku Bio-Clinical Laboratories, Inc.
  • KANAYAMA Akiko
    Chemotherapy Division, Mitsubishi Kagaku Bio-Clinical Laboratories, Inc.
  • KOBAYASHI Intetsu
    Chemotherapy Division, Mitsubishi Kagaku Bio-Clinical Laboratories, Inc.

Bibliographic Information

Other Title
  • 環境水からのLoopampレジオネラ検出試薬キットを用いたLegionella属菌の検出
  • カンキョウスイ カラノ Loopamp レジオネラ ケンシュツ シヤク キット オ モチイタ Legionellaゾクキン ノ ケンシュツ

Search this article

Description

Detection of Legionella species using the loop-mediated isothermal amplification (LAMP) method was investigated with the “Loopamp Legionella detection reagents kit E” (Eiken Chemical Co., Ltd.). The detection limit of the LAMP method was 3.7×101 CFU/test, approximately equivalent to the limit claimed by the instructions (60 CFU/test). The LAMP method detected positive reactions for 17 of 100 Legionella samples which gave negative reactions by the culture method, indicating that the sensitivity of the LAMP method was higher than the culture method. No false positive results were obtained in the screening of 5 organisms other than Legionella species. All visual results were equal to those obtained using a turbidity meter. The LAMP method can specifically detect Legionella species at levels lower than the detection limit of the conventional culture method in a short time, so is useful for the prevention of nosocomial Legionella infection.