Limitted functions of ventricular assist heart with an atrial withdrawal method.
-
- FUKUMASU H.
- Takeda Hospital
-
- YAMAZATO A.
- Takeda Hospital
-
- FUJIWARA Y.
- 京都大学医学部心臓血管外科
-
- SONEDA J.
- 京都大学医学部心臓血管外科
-
- OKAMOTO Y.
- 京都大学医学部心臓血管外科
-
- BAN T.
- 京都大学医学部心臓血管外科
-
- HIROSE H.
- 大津赤十字病院心臓血管外科
-
- YUASA S.
- 東マス技研
Bibliographic Information
- Other Title
-
- 左房脱血補助人工心臓の限界
Search this article
Description
Pulsatile ventricular assist devices have successfully been in clinical use according to the Pierce's criteria. However, there are many cases with more critical cardiac functions. Atrial withdrawal cannula can not eliminate blood flow into ventricle. So ventricular myocardium can be minimized volume load but can not be reduced pressure after-loads. Subendocardial infarction was found at autopsy following transatiral VAD in the animal heart criated more critical myocardial disfunction.