人工心肺装置のシステムによる比較  気泡型肺とローラーポンプ,膜型肺と遠心ポンプ及び気泡型肺と拍動流ローラーポンプの比較検討

書誌事項

タイトル別名
  • Comparison of three system of heart lung machine.
  • ―気泡型肺とローラーポンプ, 膜型肺と遠心ポンプ及び気泡型肺と拍動流ローラーポンプの比較検討―

この論文をさがす

説明

In cardiac operation, bubble oxygenater (Shiley S100A) and roller pump (group 1), membrane oxygenater (Capioxl II) and centrifugal pump (Bio pump)(group 2) and Shiley S100A and pulsatile roller pump (group 3) were respectively used in 9 patients, 12 patients and 10 patients. There was no significant diference about renal function between group of pulsatile flow and nonpnlsatile flow. The value of plasma free hemoglobin (Hb) concentration of group 2 and 3 were higher than that of group 1. Group 2 was divided into two groups. One was group 2-a which occurred hematuria in the early time of cardiopulmonary bypass (CPB) and the other was group 2-b. The value of free Hb concentration of group 2-a was high, compared with that of group 2-b. Perfusion pressure during CPB of group 2-a was higher than that of group 2-b. It was suggested that there was no diference among three systems of heart lung machine for usual open heart surgery. On using system of heart lung machine of group 2, α-blockade had better be administrated because of the prevention of hemolysis.

収録刊行物

  • 人工臓器

    人工臓器 16 (1), 621-624, 1987

    一般社団法人 日本人工臓器学会