Evaluation of specificity of the modified urinary ketone body test strip
-
- MATSUKI Miki
- Department of Clinical Laboratory, Juntendo University Hospital
-
- TAKEMURA Hiroyuki
- Department of Clinical Laboratory, Juntendo University Hospital
-
- UENO Tsuyoshi
- Department of Clinical Laboratory, Juntendo University Hospital
-
- WAKITA Mitsuru
- Department of Clinical Laboratory, Juntendo University Hospital
-
- KUNO Yutaka
- Department of Clinical Laboratory, Juntendo University Hospital
-
- HORII Takashi
- Department of Clinical Laboratory, Juntendo University Hospital
-
- TABE Yoko
- Department of Clinical Laboratory Medicine, Juntendo University School of Medicine
-
- OHSAKA Akimichi
- Department of Clinical Laboratory, Juntendo University Hospital
Bibliographic Information
- Other Title
-
- 尿ケトン体改良試験紙法の特異性に関する評価
Search this article
Description
To test for urinary ketone bodies, test strips based on sodium nitroprusside reaction are routinely used. However, the sodium nitroprusside reaction method frequently shows a false positive result owing to a nonspecific reaction with sulfhydryl (SH)-containing reagents including Bucillamine, an antirheumatic drug. Recently, the modified UROPAPER ALPHA 3 “Eiken” ketone test stripTM (Eiken Chemical, Japan), a urinary test strip that reduces nonspecific reactions, has been developed. In this study, the clinical usefulness of this modified urinary ketone body test strip was evaluated by comparing it with the conventional test strip UROPAPER ALPHA 3 “Eiken” test stripTM using 292 clinical samples from patients. The concordance rate of these two methods was 89.4% (261/292), and there were 31 (10.6%) unmatched samples. Twenty-five of the 31 unmatched samples showed negative results in the modified UROPAPER ALPHA 3 test and positive results in the conventional UROPAPER ALPHA 3 test. The 31 unmatched samples were finally confirmed to be negative for ketone bodies by an enzyme assay. The unmatched samples were obtained from RA patients treated with Bucillamine. In conclusion, the modified UROPAPER ALPHA 3 “Eiken” ketone test stripTM effectively reduced nonspecific reactions caused by the SH-group drug Bucillamine, enhancing the effectiveness of routine urinary tests by increasing the specificity of detecting ketone bodies.
Journal
-
- Japanese Journal of Medical Technology
-
Japanese Journal of Medical Technology 63 (5), 586-589, 2014
Japanese Association of Medical Technologists