Determination of Theophylline in Biological Fluids by High-Performance Liquid Chromatography

Bibliographic Information

Other Title
  • 高速液体クロマトグラフ法による体液中のテオフィリンの定量
  • コウソク エキタイ クロマトグラフホウ ニヨル タイエキチュウ ノ テオフィリ

Search this article

Abstract

High performance liquid chromatographic method for the determination of theophylline in biological fluids (serum, saliva and urine) was developed. Theophylline in biological fluids was adsorbed on a diatom earth column, followed by elution with a mixture of chloroform-isopropanol (95 : 5). After evaporation of the solvent, the residue was dissolved in water, injected onto a HPLC column packed with an ODS stationary phase eluted at 40°C with a mixture of 0.1M phosphate buffer (pH 5.3)-acetonitrile (9 : 1) and detected at 273 nm. 8-Chlorotheophylline was used as an internal standard. The recovery of theophylline from serum by this method was 100.1%±1.76% (SD ; n=5), Other xanthine derivatives (theobromine and caffeine) were seperated under the same conditions.

Journal

  • YAKUGAKU ZASSHI

    YAKUGAKU ZASSHI 101 (10), 955-959, 1981

    The Pharmaceutical Society of Japan

Citations (1)*help

See more

Details 詳細情報について

Report a problem