Calibration of a Superconducting Quantum Interference Device Magnetometer
-
- Koyano Tamotsu
- Cryogenics Center, University of Tsukuba
この論文をさがす
抄録
Calibration factor for magnetization scale depends on sample size along the scan direction in a superconducting quantum interference device (SQUID) magnetometer. A reference material with the same shape and size as a sample or the proper correction of apparent values is necessary for the accurate determination of magnetization.
収録刊行物
-
- Japanese Journal of Applied Physics
-
Japanese Journal of Applied Physics 43 (10), 7322-7323, 2004