CT-guided Needle Insertion Robot System “IRIS”:
-
- SEKINE Kai
- Waseda University
-
- TSUMURA Ryosuke
- Waseda University
-
- IKEDA Iori
- Waseda University
-
- IWATA Hiroyasu
- Waseda University
Bibliographic Information
- Other Title
-
- CTガイド下穿刺支援ロボットIRISの開発
- - Feasibility study of total system-
- ―統合システムにおけるフィージビリティスタディ―
Abstract
<p>In recent years, HITV therapy, which cures cancer by improving the immunity of patients, has attracted attention. In this therapy, high puncture skills are required for physicians because dendritic cells are injected directly into the tumor using a fine needle. In order to solve this problem, we developed a CT-guided needle insertion robot “IRIS” that provides accurate puncture of a fine needle. In this study, we evaluated the puncture accuracy using an abdominal Biopsy phantom as a feasibility study of the total system. As a result, the puncture accuracy was less than the required accuracy ± 3 mm, suggesting the usefulness of IRIS for practical use.</p>
Journal
-
- The Proceedings of JSME annual Conference on Robotics and Mechatronics (Robomec)
-
The Proceedings of JSME annual Conference on Robotics and Mechatronics (Robomec) 2020 (0), 2A1-E05-, 2020
The Japan Society of Mechanical Engineers
- Tweet
Details 詳細情報について
-
- CRID
- 1390286426521772416
-
- NII Article ID
- 130007943910
-
- ISSN
- 24243124
-
- Text Lang
- ja
-
- Data Source
-
- JaLC
- Crossref
- CiNii Articles
-
- Abstract License Flag
- Disallowed