書誌事項
- タイトル別名
-
- Proposal of Magnetically Driven Biopsy Forceps for Medical Capsule
抄録
<p>This paper has proposed a swallowable medical capsule with a magnetically driven biopsy forceps. The biopsy unit is composed of a biopsy forceps and a lead screw actuator with a permanent magnet. When a rotating magnetic field is externally applied, the forceps is projected from a front of the capsule by the lead screw actuator. When the magnetic field is further continuously applied, the forceps is opened at a lesion. Then, by applying the rotating magnetic field reversely, the forceps is closed to collect the tissue and withdrawn into the capsule. This report describes the basic structure, actuation principle and successful operation results by the prototype.</p>
収録刊行物
-
- 電気学会論文誌. A
-
電気学会論文誌. A 142 (6), 243-249, 2022-06-01
一般社団法人 電気学会
- Tweet
詳細情報 詳細情報について
-
- CRID
- 1390855201291985920
-
- ISSN
- 13475533
- 03854205
-
- 本文言語コード
- ja
-
- データソース種別
-
- JaLC
- Crossref
- KAKEN
-
- 抄録ライセンスフラグ
- 使用不可